Ultra-low foetal radiation exposure in 18F-FDG PET/CT imaging with a long axial field-of-view PET/CT system
作者:Charlotte L. C. Smith, M. Yaqub, R. Wellenberg, J. J. Knip, R. Boellaard, G. Zwezerijnen · 发表于:EJNMMI Physics · 年份:2024 · DOI:10.1186/s40658-024-00648-w · 被引用次数:14 · 研究领域:Medicine
Long axial field-of-view (LAFOV) PET/CT systems enable PET/CT scans with reduced injected activities because of improved sensitivity. With this study, we aimed to examine the foetal radiation dose from an 18F-FDG PET/CT scan on a LAFOV PET/CT system with reduced injected activity. Two pregnant women were retrospectively included and received an 18F-FDG PET/CT scan on a LAFOV PET/CT system with an intravenous bolus injection of 0.30 MBq/kg. Foetal radiation exposure from the PET was estimated using dose conversion factors from three published papers. Radiation exposure from the CT scans was estimated using CT-Expo. Foetal radiation dose from the PET scan ranged between 0.11 and 0.44 mGy. Foetal radiation exposure from the CT scan ranged between < 0.10 – 0.90 mGy depending if the foetus was included in the field-of-view. Foetal radiation dose could be reduced to < 1.5 mGy when scanning pregnant patients on a LAFOV PET/CT system. The radiation dose to the foetus was reduced significantly in our study due to the increased sensitivity of the LAFOV PET/CT system.