Scholay

学术搜索 · AI 审稿 · LaTeX 协作

Noise and artifact suppression in SQUID and wearable OPM-MEG: A systematic review of background, physiological, and Technical interference

作者:Ruonan Wang, MA Yu-jie, Ruochen Zhao, Jin Ding, Ling Li, Yanfei Yang, Fulong Wang, Zhiqiang Cao, Xueying Zhang, Xiaoyang Lin, Xiaolin Ning · 发表于:NeuroImage · 年份:2025 · DOI:10.1016/j.neuroimage.2025.121403 · 被引用次数:7 · 研究领域:Atomic and Subatomic Physics Research、Non-Invasive Vital Sign Monitoring、Functional Brain Connectivity Studies

Magnetoencephalography (MEG) is a non-invasive imaging technique that captures neural activity with high spatio-temporal resolution. In recent years, novel wearable devices based on Optically Pumped Magnetometer (OPM) have emerged as a new driving force for advancing MEG due to their cost-effectiveness, portability, and mobility. In practical applications, MEG signals are frequently influenced by various interference sources, resulting in degradation of signal quality. Consequently, numerous suppression techniques have been proposed to overcome these challenges. This manuscript presents a comprehensive review of the most advanced methods for suppressing MEG noise or artifacts, with a specific focus on mitigating background noise, physiological artifacts (such as those caused by heartbeat, eye movements, and muscle contractions), as well as technical artifacts (including system-related artifacts associated with devices, motion-induced artifacts, and metal-induced artifacts). Additionally, the current limitations and challenges of these approaches in real-world scenarios are highlighted. Reviewing nearly a decade of research, there is an urgent need for a lightweight noise analysis framework in the complex measurement environment of wearable OPM-MEG devices. This framework should be capable of effectively detecting, classifying, and suppressing individual and combined MEG interference. By addressing this need, we can enhance the reliability and practicality of MEG signals while...