A Sandwich-Type Impedimetric Immunosensor for the Detection of Tau-441 Biomarker
作者:Khouloud Djebbi, Yang Xiang, Biao Shi, Lyes Douadji, Xiaohan Chen, Jing Liu, Chaker Tlili, Deqiang Wang · 发表于:Bioengineering · 年份:2025 · DOI:10.3390/bioengineering12080805 · 被引用次数:4 · 研究领域:Advanced biosensing and bioanalysis techniques、Molecular Junctions and Nanostructures、Electrochemical sensors and biosensors
The human Tau protein stands for one of the most conspicuous and crucial hallmarks of Alzheimer's disease (AD) diagnosis, along with other tauopathies. However, the assay for direct detection of tiny Tau protein concentrations in human samples continues to pose a significant challenge for the early diagnosis of AD. Thus, an amplification-based strategy is required. In this proposed work, we established an impedimetric immunosensor to detect human Tau-441 protein in PBS buffer using a sandwich approach, wherein we employed two distinct monoclonal antibodies (HT7 and BT2) that specifically recognize the amino acids 159-198 of the target protein. Through this strategy, we were able to detect as low as 0.08 pg/mL. These findings were attributed to the use of a biotinylated antibody (BT2)-streptavidin complex, which facilitated the amplification of the normalized signal, resulting in a lower limit of detection in comparison to the directly based immunosensors. Subsequently, we investigated the designed immunosensor to assess the assay's selectivity in the presence of different off-targets, and no cross-interaction was recorded. The outcomes of our study provide valuable new insights into the application of sandwich-based assay as a highly sensitive and selective immunosensor for the detection of small protein.