Miniature Mass Spectrometry for Point-of-Care Testing the GPIIb/IIIa Inhibitor Tirofiban during Perioperative Period of Percutaneous Coronary Intervention

作者:Dongchen Zhou, Jiahui Wu, Qingcheng Wang, Yong Liu, Shiqi Wang, Weizong Zhang, Yunfeng Zhang, Huizhi Ding, Yunfei Shao, Haixing Wang, Qing Shen · 发表于:ACS Omega · 年份:2024 · DOI:10.1021/acsomega.4c06581 · 被引用次数:5 · 研究领域:Acute Myocardial Infarction Research、Heparin-Induced Thrombocytopenia and Thrombosis、Platelet Disorders and Treatments

High Resolution Image Download MS PowerPoint Slide Precise administration of tirofiban must be carefully considered to achieve the best treatment efficacy and maximize safety for patients. Herein, a paper spray ionization (PSI) linear ion trap (LIT) portable mass spectrometer (pMS) based point-of-care testing (POCT) technique was developed for on-site sampling, clinical testing, and immediate analysis of tirofiban blood drug concentrations. The results showed that tirofiban formed a significant and stable parent ion peak at m / z 441.3 by MS1 full scan in positive ion mode, which fragmented into product ions at m / z 395.4, m / z 321.3, m / z 276.3, and m / z 260.3 through collision-induced dissociation (CID) in MS/MS. To improve ion response, the parameters were optimized to the voltages of ionization 3600 V, ion isolation 1 (ISO1) 8 V, ion isolation 2 (ISO2) 2 V, and collision-induced dissociation (CID) 3 V. This method was validated, and the limit of detection (LOD) and limit of quantification (LOQ) were found to be 10.1 and 33.7 μg·L –1, respectively. For precision, it had the relative standard deviation (RSD) of interday precision of 4.8 to 6.7% and the RSD of intraday precision of 7.8 to 8.3%. The recovery of the method ranged from 87.5 to 93.4%. Although matrix effects in blood samples had some inhibitory effects on the target signal formation, the method compensated for part of the matrix effects by establishing a matrix-matched calibration curve, which exhibited good...