A comparison study of different eddy current measurements at low field portable MRI

作者:Suen Chen, Yunhao Xie, Yueqi Qiu, Hao Chen, Haijun Luo, Zhiyong Zhang · 发表于:Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition · 年份:2024 · DOI:10.58530/2024/4077 · 被引用次数:1 · 研究领域:Non-Destructive Testing Techniques、Nuclear reactor physics and engineering、Welding Techniques and Residual Stresses

Motivation: Eddy current results in resolution degradation and unwanted phase variations. However, there have been relatively few eddy-current related studies carried out in low field recently. Goal(s): Our goal was to introduce and compare image-phase based and FID-phase based eddy current measurement techniques in a 64mT MRI system and reduced eddy-current effects. Approach: We employed a large phantom or six field probes to measure eddy-current-induced gradients shift. And we modeled them as sums of exponentially decaying components to compensate the gradient unit. Results: Eddy-current-induced field shift was initially about 350ppm, and reduced to 240ppm and 20ppm by these two methods. Impact: This work develops two methods for measuring eddy current in low-field MRI, especially for portable scanner, which help reducing eddy-current-induced effects and improving the image quality of clinical sequence (e.g. DWI, DTI) in low field scanner.