Effect of varying cuff sizes with identical inner diameter on endotracheal intubation in critically ill adults: A sealed tracheal controlled trial

作者:Yannan Zhang, Hai-Yan Shi, Wang‐Qin Shen, Jiahai Shi, Yanping Zhu, Yanghui Xu, Honglei Wu · 发表于:Medicine · 年份:2024 · DOI:10.1097/md.0000000000038326 · 被引用次数:3 · 研究领域:Airway Management and Intubation Techniques、Nosocomial Infections in ICU、Simulation-Based Education in Healthcare

BACKGROUND: The present study aims to determine the impact of different cuff diameters on the cuff pressure of endotracheal tubes (ETTs) when the trachea is adequately sealed. METHODS: In the present single-center clinical trial, adult patients who underwent cardiothoracic surgery were assigned to use ETTs from 2 brands (GME and GZW). The primary endpoint comprised of the following: cuff diameter, inner diameter of the ETT, manufacturer, and the number of subjects with tracheal leakage when the cuff pressure was 30 cm H2O. RESULTS: A total of 298 patients were assigned into 2 groups, based on the 2 distinct brands of ETTs: experimental group (n = 122, GME brand) and control group (n = 176, GZW brand). There were no significant differences in baseline characteristics. However, the cuff diameter was significantly smaller in the control group, when compared to the experimental group (P = .001), and the incidence of tracheal leakage was significantly higher in the control group (P = .001). Furthermore, the GME brand ETT had a significantly larger cuff diameter, when compared to the GZW brand ETT. CONCLUSION: The cuff size would mismatch the tracheal area in clinical practice. Therefore, chest computed tomography is recommended to routinely evaluate the tracheal cross-sectional area during anesthesia, in order to ensure the appropriate cuff size selection.