Tricuspid valve flow measurement using a deep learning framework for automated valve‐tracking 2D phase contrast

作者:Jérôme Lamy, Ricardo A. Gonzales, Jie Xiang, Felicia Seemann, Steffen Huber, Jeremy Steele, Oliver Wieben, Einar Heiberg, Dana C. Peters · 发表于:Magnetic Resonance in Medicine · 年份:2024 · DOI:10.1002/mrm.30163 · 被引用次数:6 · 研究领域:Cardiac Valve Diseases and Treatments、Cardiovascular Function and Risk Factors、Pulmonary Hypertension Research and Treatments

PURPOSE: Tricuspid valve flow velocities are challenging to measure with cardiovascular MR, as the rapidly moving valvular plane prohibits direct flow evaluation, but they are vitally important to diastolic function evaluation. We developed an automated valve-tracking 2D method for measuring flow through the dynamic tricuspid valve. METHODS: Nine healthy subjects and 2 patients were imaged. The approach uses a previously trained deep learning network, TVnet, to automatically track the tricuspid valve plane from long-axis cine images. Subsequently, the tracking information is used to acquire 2D phase contrast (PC) with a dynamic (moving) acquisition plane that tracks the valve. Direct diastolic net flows evaluated from the dynamic PC sequence were compared with flows from 2D-PC scans acquired in a static slice localized at the end-systolic valve position, and also ventricular stroke volumes (SVs) using both planimetry and 2D PC of the great vessels. RESULTS: The mean tricuspid valve systolic excursion was 17.8 ± 2.5 mm. The 2D valve-tracking PC net diastolic flow showed excellent correlation with SV by right-ventricle planimetry (bias ± 1.96 SD = -0.2 ± 10.4 mL, intraclass correlation coefficient [ICC] = 0.92) and aortic PC (-1.0 ± 13.8 mL, ICC = 0.87). In comparison, static tricuspid valve 2D PC also showed a strong correlation but had greater bias (p = 0.01) versus the right-ventricle SV (10.6 ± 16.1 mL, ICC = 0.61). In most (8 of 9) healthy subjects, trace regurgitation was...