Scholay

学术搜索 · AI 审稿 · LaTeX 协作

The 3-Dimensional Intelligent Structured Light Technique: A New Registration Method in Stereotactic Neurosurgery

作者:Du Cai, Xiu Wang, Wenhan Hu, Jiajie Mo, Huanguang Liu, Xiaoyan Li, Xixi Zheng, Xiaosheng Ding, Juan An, Yichun Hua, Jianguo Zhang, Kai Zhang, Chao Zhang · 发表于:Operative Neurosurgery · 年份:2024 · DOI:10.1227/ons.0000000000001184 · 被引用次数:7 · 研究领域:Soft Robotics and Applications、Advanced Radiotherapy Techniques、Facial Nerve Paralysis Treatment and Research

BACKGROUND AND OBJECTIVES: Surface-based facial scanning registration emerged as an essential registration method in the robot-assisted neuronavigation surgery, providing a marker-free way to align a patient's facial surface with the imaging data. The 3-dimensional (3D) structured light was developed as an advanced registration method based on surface-based facial scanning registration. We aspire to introduce the 3D structured light as a new registration method in the procedure of the robot-assisted neurosurgery and assess the accuracy, efficiency, and safety of this method by analyzing the relative operative results. METHODS: We analyzed the results of 47 patients who underwent Ommaya reservoir implantation (n = 17) and stereotactic biopsy (n = 30) assisted by 3D structured light at our hospital from January 2022 to May 2023. The accuracy and additional operative results were analyzed. RESULTS: For the Ommaya reservoir implantation, the target point error was 3.2 ± 2.2 mm and the entry point error was 3.3 ± 2.4 mm, while the operation duration was 35.8 ± 8.3 minutes. For the stereotactic biopsy, the target point error was 2.3 ± 1.3 mm and the entry point error was 2.7 ± 1.2 mm, while the operation duration was 24.5 ± 6.3 minutes. CONCLUSION: The 3D structured light technique reduces the patients' discomfort and offers the advantage of a simpler procedure, which can improve the clinical efficiency with the sufficient accuracy and safety to meet the clinical requirements of th...