Novel Technology Allowing Cone Beam Computed Tomography in 6 Seconds: A Patient Study of Comparative Image Quality
作者:James L. Robar, Amanda Cherpak, R. Lee MacDonald, Abigail Yashayaeva, David McAloney, Natasha McMaster, Kenny Zhan, Slawa Cwajna, Nikhilesh Patil, Hannah Dahn · 发表于:Practical Radiation Oncology · 年份:2023 · DOI:10.1016/j.prro.2023.10.014 · 被引用次数:101 · 研究领域:Medical Imaging Techniques and Applications、Advanced X-ray and CT Imaging、Radiation Dose and Imaging
PURPOSE: The goal of this study was to evaluate the image quality provided by a novel cone beam computed tomography (CBCT) platform (HyperSight, Varian Medical Systems), a platform with enhanced reconstruction algorithms as well as rapid acquisition times. Image quality was compared with both status quo CBCT for image guidance, and to fan beam CT (FBCT) acquired on a CT simulator (CTsim). METHODS AND MATERIALS: In a clinical study, 30 individuals were recruited for whom either deep inspiration (DIBH) or deep exhalation breath hold (DEBH) was used during imaging and radiation treatment of tumors involving liver, lung, breast, abdomen, chest wall, and pancreatic sites. All subjects were imaged during breath hold with CBCT on a standard image guidance platform (TrueBeam 2.7, Varian Medical Systems) and FBCT CT (CTsim, GE Optima). HyperSight imaging with both breath hold (HSBH) and free breathing (HSFB) was performed in a single session. The 4 image sets thus acquired were registered and compared using metrics quantifying artifact index, image nonuniformity, contrast, contrast-to-noise ratio, and difference of Hounsfield unit (HU) from CTsim. RESULTS: HSBH provided less severe artifacts compared with both HSFB and TrueBeam. The severity of artifacts in HSBH images was similar to that in CTsim images, with statistically similar artifact index values. CTsim provided the best image uniformity; however, HSBH provided improved uniformity compared with both HSFB and TrueBeam. CTsim dem...