Influence of a Capsular Tension Ring on Capsular Bag Behavior of a Plate Haptic Intraocular Lens

作者:Daniel Schartmüller, Veronika Röggla, Luca Schwarzenbacher, Elias Laurin Meyer, Claudette Abela‐Formanek, Christina Leydolt, Rupert Menapace · 发表于:Ophthalmology · 年份:2023 · DOI:10.1016/j.ophtha.2023.10.031 · 被引用次数:20 · 研究领域:Intraocular Surgery and Lenses、Ophthalmology and Visual Impairment Studies、Retinal and Macular Surgery

PURPOSE: To evaluate the influence of a capsular tension ring (CTR) on rotational stability, decentration, tilt, and axial stability of an 11.0-mm plate haptic intraocular lens (IOL). DESIGN: Intraindividual, randomized, double-masked, controlled clinical trial. PARTICIPANTS: Patients scheduled for sequential same-day bilateral cataract surgery. METHODS: All patients were randomized to receive a CTR and a plate haptic IOL in one eye and a plate haptic IOL in the fellow eye only. Intraocular lens axis assessment was performed at the end of surgery, 1 hour, 1 week, 1 month, and 6 months using a high-precision evaluation method. Decentration and tilt of the crystalline and pseudophakic lenses were assessed before surgery and at 1 week and 6 months using an anterior segment OCT. MAIN OUTCOME MEASURES: Rotational stability from the end of surgery to 6 months and at all follow-up visits, decentration and tilt at 6 months, and differences in axial shift between 1 week and 6 months. RESULTS: One hundred thirty eyes of 65 patients were included in the study. Absolute rotation from the end of surgery to 6 months was 2.8 ± 3.9° and 3.2 ± 5.3° for the CTR and control groups, respectively (P = 0.613). Intraocular lens decentration and IOL tilt at 6 months were 0.29 ± 0.1 mm and 0.24 ± 0.1 mm and 6.7 ± 2.8° and 5.6 ± 1.6° for the CTR and control groups, respectively (P = 0.058; P < 0.01). A posterior IOL shift of 0.31 ± 0.31 mm and 0.19 ± 0.14 mm was observed in the CTR and control groups,...