Scholay

学术搜索 · AI 审稿 · LaTeX 协作

Wireless, Smart Hemostasis Device with All‐Soft Sensing System for Quantitative and Real‐Time Pressure Evaluation

作者:Chengjun Zhang, Qing Yang, Xianglin Meng, Haoyu Li, Zexiang Luo, Lin Kai, Jie Liang, Sicheng Chen, Feng Chen · 发表于:Advanced Science · 年份:2023 · DOI:10.1002/advs.202303418 · 被引用次数:33 · 研究领域:Advanced Sensor and Energy Harvesting Materials、Advanced Fiber Optic Sensors、Non-Invasive Vital Sign Monitoring

Abstract The properly applied pressure between the skin and hemostasis devices is an essential parameter for preventing bleeding and postoperative complications after a transradial procedure. However, this parameter is usually controlled based on the subjective judgment of doctors, which might cause insufficient hemostatic effect or thrombosis. Here this study develops a compact and wireless sensing system for continuously monitoring the pressure applied on the radial artery and wrist skin in clinical practice. A liquid metal (LM)‐based all‐soft pressure sensor is fabricated to enable conformal attachment between the device and skin even under large deformation conditions. The linear sensitivity of 0.007 kPa −1 among the wide pressure range of 0–100 kPa is achieved and the real‐time detection data can be wirelessly transmitted to mobile clients as a reference pressure value. With these devices, detailed pressure data can be collected, analyzed, and stored for medical assistance as well as to improve surgery quality.