Validation of the relative biological effectiveness of active-energy scanning carbon-ion radiotherapy on a commercial treatment planning system with a microdosimetic kinetic model
作者:Weiwei Wang, Wei Sun, Hao Shen, Jingfang Zhao · 发表于:Radiation Oncology · 年份:2023 · DOI:10.1186/s13014-023-02267-8 · 被引用次数:5 · 研究领域:Radiation Therapy and Dosimetry、Advanced Radiotherapy Techniques、Effects of Radiation Exposure
BACKGROUND: The study objective was to validate the relative biological effectiveness (RBE) calculated by the modified microdosimetric kinetic model in RayStation (Ray-MKM) for active-energy scanning carbon-ion radiotherapy. METHODS: The Ray-MKM was benchmarked using a spread-out Bragg-peak (SOBP) plan, which was suggested in literature from the National Institute of Radiobiological Science (NIRS) in Japan. The residual RBE differences from the MKM at NIRS (NIRS-MKM) were derived using several SOBP plans with different ranges, SOBP widths, and prescriptions. To investigate the origins of the differences, we compared the saturation-corrected dose-mean specific energy [Formula: see text] of the aforementioned SOBPs. Furthermore, we converted the RBE-weighted doses with the Ray-MKM to those with local effect model I (LEM doses). The purpose was to investigate whether the Ray-MKM could reproduce the RBE-weighted conversion study. RESULTS: The benchmark determined the value of the clinical dose scaling factor, [Formula: see text], as 2.40. The target mean RBE deviations between the Ray-MKM and NIRS-MKM were median: 0.6 (minimum: 0.0 to maximum: 1.69) %. The [Formula: see text] difference in-depth led to the RBE difference in-depth and was remarkable at the distal end. The converted LEM doses from the Ray-MKM doses were comparable (the deviation being - 1.8-0.7%) to existing literature. CONCLUSION: This study validated the Ray-MKM based on our active-energy scanning carbon-ion beam...