Self-Measured Doppler Blood Pressure Readings in Continuous-Flow LVAD Patients are Reliable
作者:D. Robson, Audrey Adji, S. Conte, Phillip Lo, S. Emmanuel, Naomi M. Anderson, Maria D. Correas, K. Muthiah, Chris Hayward · 发表于:ASAIO Journal · 年份:2022 · DOI:10.1097/01.mat.0000840724.82147.05 · 被引用次数:5 · 研究领域:Mechanical Circulatory Support Devices、Respiratory Support and Mechanisms、Cardiac Arrest and Resuscitation
Background: Blood pressure (BP) management is imperative in reducing risk of adverse events in continuous flow (CF) left ventricular assist device (LVAD) patients. The gold standard for non-invasive blood pressure (BP) measurement in these patients is Doppler ultrasound, which requires training and dexterity and is not routinely used in the outpatient setting. In the current COVID-19 pandemic era, telehealth assessment has become mandatory to assess and monitor patients. At our institution, we have issued all CF-LVAD patients with a Doppler ultrasound device and an oscillometric BP monitor, and provided training to correctly use these devices. Accuracy of BP measurements in this setting has not been confirmed. Aim: We examined whether Doppler BP obtained by CF-LVAD patients (pts) correlates to Doppler BP obtained by health care workers (HCWs). Methods: CF-LVAD pts were recruited between November 2020 and January 2022. During these routine outpatient follow-up visits, Doppler BP measurements were obtained simultaneously and independently by pts and HCW. BP was taken in a comfortable seating position following 5 minutes rest. Results: A total 223 pairs of simultaneous patient Doppler BP readings and HCW Doppler BP readings were obtained from 38 CF-LVAD patients (87% male, mean age at implant 52 + 14 years, 15 HeartMate3 and 23 HeartWare HVAD, with support duration range 28 – 3492 days). Patient Doppler BP (average 85.9 + 9.8 mmHg) correlated closely with HCW Doppler BP (average...