The influence of non‐conformal grid interfaces on the results of large eddy simulation of centrifugal blood pumps

作者:Peng Wu, Jia‐Dong Huo, Zijian Zhang, Chunju Wang · 发表于:Artificial Organs · 年份:2022 · DOI:10.1111/aor.14263 · 被引用次数:15 · 研究领域:Mechanical Circulatory Support Devices、Intravenous Infusion Technology and Safety、Blood properties and coagulation

BACKGROUND: Computational fluid dynamics has been widely used to assist the design and evaluation of blood pumps. Discretization errors associated with computational grid may influence the credibility of numerical simulations. Non-conformal grid interfaces commonly exist in rotary machines, including rotary blood pumps. Should grid size across the interface differ greatly, large errors may occur. METHODS: This study explored the effects of non-conformal grid interface on the prediction of the flow field and hemolysis in blood pumps using large eddy simulation (LES). Two benchmarks, a nozzle model and a centrifugal blood pump were chosen as test cases. RESULTS: This study found that non-conformal grid interfaces with considerable change of grid sizes led to discontinuities of flow variables and brought errors to metrics such as pressure head (7%) and hemolysis (up to 14%). CONCLUSIONS: The results on the full unstructured grid are more accurate with negligible changes of flow variables across the non-conformal grid interface. A full unstructured grid should be employed for centrifugal blood pumps to minimize the influence of non-conformal grid interfaces for LES simulations.