Evaluation of a single-breath-hold radial turbo-spin-echo sequence for T2 mapping of the liver at 3T

作者:Diana Benčíková, Fei Han, Stephan Kannengieser, Marcus Raudner, Sarah Poetter‐Lang, Nina Bastati, Gert Reiter, Raphael Ambros, Ahmed Ba‐Ssalamah, Siegfried Trattnig, Martin Krššák · 发表于:European Radiology · 年份:2021 · DOI:10.1007/s00330-021-08439-y · 被引用次数:13 · 研究领域:Advanced MRI Techniques and Applications、MRI in cancer diagnosis、Advanced NMR Techniques and Applications

Abstract Objectives T2 mapping of the liver is a potential diagnostic tool, but conventional techniques are difficult to perform in clinical practice due to long scan time. We aimed to evaluate the accuracy of a prototype radial turbo-spin-echo (rTSE) sequence, optimized for multi-slice T2 mapping in the abdomen during one breath-hold at 3 T. Methods A multi-sample (fat: 0–35%) agarose phantom doped with MnCl 2 and 80 subjects (73 patients undergoing abdomen MR examination and 7 healthy volunteers) were investigated. A radial turbo-spin-echo (rTSE) sequence with and without fat suppression, a Cartesian turbo-spin-echo (Cart-TSE) sequence, and a single-voxel multi-echo STEAM spectroscopy (HISTO) were performed in phantom, and fat-suppressed rTSE and HISTO sequences were performed in in vivo measurements. Two approaches were used to sample T2 values: manually selected circular ROIs and whole liver analysis with Gaussian mixture models (GMM). Results The rTSE-T2s values exhibited a strong correlation with Cart-TSE-T2s ( R 2 = 0.988) and with HISTO-T2s of water ( R 2 = 0.972) in phantom with an offset between rTSE and Cart-TSE maps (mean difference = 3.17 ± 1.18 ms). The application of fat suppression decreased T2 values, and the effect was directly proportional to the amount of fat. Measurements in patients yielded a linear relationship between rTSE- and HISTO-T2s ( R 2 = 0.546 and R 2 = 0.580 for ROI and GMM, respectively). Conclusion The fat-suppressed rTSE sequence allows for...