Competitive Real-Time Near Infrared (NIR) Vein Finder Imaging Device to Improve Peripheral Subcutaneous Vein Selection in Venipuncture for Clinical Laboratory Testing
作者:Mark D. Francisco, Wen‐Fan Chen, Cheng-Tang Pan, Ming‐Cheng Lin, Zhi‐Hong Wen, Chien-Feng Liao, Yow‐Ling Shiue · 发表于:Micromachines · 年份:2021 · DOI:10.3390/mi12040373 · 被引用次数:46 · 研究领域:Reconstructive Surgery and Microvascular Techniques、Diagnosis and Treatment of Venous Diseases、Central Venous Catheters and Hemodialysis
In this study, near-infrared (NIR) technology was utilized to develop a low-cost real-time near infrared (NIR) guiding device for cannulation. A portable device that can be used by medical practitioners and also by students for their skills development training in performing cannulation. METHODS: First, is the development of a reflectance type optical vein finder using three (3) light emitting diode (LED) lights with 960 nm wavelength, complementary metal-oxide-semiconductor-infrared (CMOS-IR) sensor camera with 1920 × 1080 UXGA (1080P), IR filter set for the given wavelength, and an open-source image processing software. Second, is the actual in-vitro human testing in two sites: the arm and dorsal hand of 242 subjects. The following parameters were included, such as gender, age, mass index (BMI), and skin tone. In order to maximize the assessment process towards the device, the researchers included the arm circumference. This augmented subcutaneous vein imaging study using the develop vein finder device compared the difference in the captured vein images through visual and digital imaging approaches. The human testing was performed in accordance with the ethical standards of the Trinity University of Asia-Institutional Ethics Review Committee (TUA-IERC). RESULTS: The NIR imaging system of the developed vein finder in this study showed its capability as an efficient guiding device through real-time vein pattern recognition, for both sites. Improved captured vein images were o...