Associations between anterior segment parameters and rotational stability of a plate-haptic toric intraocular lens
作者:Yunqian Yao, Jiaqi Meng, Wenwen He, Keke Zhang, Ling Wei, Kaiwen Cheng, Yi Lü, Xiangjia Zhu · 发表于:Journal of Cataract & Refractive Surgery · 年份:2021 · DOI:10.1097/j.jcrs.0000000000000653 · 被引用次数:35 · 研究领域:Ophthalmology and Visual Impairment Studies、Intraocular Surgery and Lenses、Glaucoma and retinal disorders
PURPOSE: To evaluate the associations between anterior segment parameters and the rotational stability of a plate-haptic toric intraocular lens (IOL). SETTING: Eye and Ear, Nose, Throat Hospital of Fudan University. DESIGN: Retrospective case series. METHODS: Patients who underwent uneventful phacoemulsification and plate-haptic toric IOL (AT TORBI 709M IOL) implantation were included. Preoperative axial length (AL) and anterior segment parameters, including the white-to-white (WTW) distance, anterior chamber depth (ACD), lens thickness (LT), and anterior segment length (ASL; the sum of ACD and LT) were recorded. IOL rotation, residual astigmatism (RAS), and visual acuity were evaluated 1 month postoperatively. The associations between the anterior segment parameters and IOL rotation were evaluated. RESULTS: A total of 102 eyes of 102 patients were included. The mean AL was 26.43 ± 2.65 mm (range 21.71-34.60 mm). The mean IOL rotation was 4.59 ± 3.18 degrees, and RAS was 0.62 ± 0.39 D postoperatively. No correlation was detected between AL, ACD, or LT and the rotation of the plate-haptic toric IOL (all P > .05). However, its rotation correlated positively with the WTW distance (r = 0.250, P = .011) and ASL (r = 0.214, P = .030). Backward stepwise multiple linear regression revealed that the WTW distance (β = 2.142, P = .014) and ASL (β = 2.060, P = .037) were independent predictors of plate-haptic toric IOL rotation. CONCLUSIONS: Plate-haptic toric IOLs rotate more in eyes wi...