Scholay

学术搜索 · AI 审稿 · LaTeX 协作

A Novel Dual-Hormone Insulin-and-Pramlintide Artificial Pancreas for Type 1 Diabetes: A Randomized Controlled Crossover Trial

作者:Ahmad Haidar, Michael A. Tsoukas, Sarah Bernier-Twardy, Jean‐François Yale, Joanna Rutkowski, Anne Bossy, Evelyne Pytka, Anas El Fathi, Natalia Strauss, Laurent Legault · 发表于:Diabetes Care · 年份:2020 · DOI:10.2337/dc19-1922 · 被引用次数:137 · 研究领域:Diabetes Management and Research、Pancreatic function and diabetes、Diabetes and associated disorders

OBJECTIVE The rapid insulin-alone artificial pancreas improves glycemia in type 1 diabetes but daytime control remains suboptimal. We propose two novel dual-hormone artificial pancreas systems. RESEARCH DESIGN AND METHODS We conducted a randomized crossover trial comparing a rapid insulin-alone artificial pancreas with rapid insulin-and-pramlintide and with regular insulin-and-pramlintide artificial pancreas systems in adults with type 1 diabetes. Participants were assigned to the interventions in random order during three 24-h inpatient visits. Each visit was preceded by an outpatient hormonal open-loop run-in period of 10–14 days. The dual-hormone artificial pancreas delivered pramlintide in a basal-bolus manner, using a novel dosing algorithm, with a fixed ratio relative to insulin. The primary outcome was time in the range 3.9–10.0 mmol/L. RESULTS Compared with the rapid insulin-alone artificial pancreas system, the rapid insulin-and-pramlintide system increased the time in range from 74% (SD 18%) to 84% (13%) (P = 0.0014), whereas the regular insulin-and-pramlintide system did not change the time in range (69% [19%]; P = 0.22). The increased time in range with the rapid insulin-and-pramlintide system was due to improved daytime control (daytime time in range increased from 63% [23%] to 78% [16%], P = 0.0004). There were 11 (1 per 2.5 days) hypoglycemic events (<3.3 mmol/L with symptoms or <3.0 mmol/L irrespective of symptoms) with the rapid insulin-alone system, compared...