Real-Time Monitoring of Contact Impedance From Multiple Electrode–Scalp Interfaces During Cerebral Electrical Impedance Tomography

作者:Hang Ma, Haoting Li, Xuechao Liu, Weichen Li, Junying Xia, Benyuan Liu, Xuetao Shi, Xiuzhen Dong, Feng Fu · 发表于:IEEE Access · 年份:2019 · DOI:10.1109/access.2019.2928580 · 被引用次数:25 · 研究领域:Electrical and Bioimpedance Tomography、Hemodynamic Monitoring and Therapy、Microfluidic and Bio-sensing Technologies

Electrical impedance tomography (EIT) is a promising medical technique for monitoring brain injury through the reflection of electrical impedance changes in internal brain tissues. However, the contact impedances at the electrode-scalp interfaces can still notably affect the EIT accuracy. Hence, we propose a method for real-time monitoring the multi-channel contact impedances of cerebral EIT using a reference electrode attached close to the head vertex. This method allows to approximate inter-channel differences and changes over time of multi-channel contact impedances. The experimental results on 36 volunteers show that the maximum relative inter-channel difference obtained from 16 disposable surface electrodes reaches 18.7%, and the maximum relative change over time reaches 19.2%. The proposed method can assist clinicians in real-time monitoring of contact conditions from the multiple electrode-scalp interfaces during the clinical application of the cerebral EIT.