Scholay

学术搜索 · AI 审稿 · LaTeX 协作

Pptv-Level Intra-Cavity QEPAS Sensor for Acetylene Detection Using a High Power Q-Switched Fiber Laser

作者:Qinduan Zhang, Jun Chang, Zhenhua Cong, Yiwen Feng, Zongliang Wang, Fupeng Wang, Jiachen Sun · 发表于:IEEE Sensors Journal · 年份:2019 · DOI:10.1109/jsen.2019.2910665 · 被引用次数:21 · 研究领域:Spectroscopy and Laser Applications、Atmospheric Ozone and Climate、Advanced Fiber Laser Technologies

A high-sensitivity quartz enhanced photoacoustic spectroscopy (QEPAS) system based on acousto-optic Q-switched fiber laser for C2H2detection was proposed and experimentally demonstrated. The gas sensor system coupled an erbium-doped fiber laser with a custom-made transmission quartz tuning fork structure. An acousto-optic modulator was used to perform amplitude modulation at a modulation frequency of 32.76 kHz and the system achieved a maximum intra-cavity peak pulse power of 6 W. The system response relationship between the peak pulse power and QEPAS signals confirms that saturation does not occur with the laser power less than 6 W. Preliminary measurements of the sensor response to C2H2concentrations have been made with pure N2as a carrier gas, and the experimental results showed the relationship between QEPAS signal amplitude and the C2H2concentrations have a good linearity (R-square = 0.99918). Finally, trace gas measurements have been assessed and a minimum detection limit of 101 pptv within 20 ms integration time for C2H2has been achieved. This paper provides a valuable gas sensor for practical project application.