Direct intensity calibration of X-ray grazing-incidence microscopes with home-lab source

作者:Yaran Li, Qing Xie, Zhiqiang Chen, Qiuqi Xin, Xin Wang, Baozhong Mu, Zhanshan Wang, Shenye Liu, Yongkun Ding · 发表于:Review of Scientific Instruments · 年份:2018 · DOI:10.1063/1.5003959 · 被引用次数:6 · 研究领域:Laser-Plasma Interactions and Diagnostics、Advanced X-ray Imaging Techniques、High-pressure geophysics and materials

Direct intensity calibration of X-ray grazing-incidence microscopes is urgently needed in quantitative studies of X-ray emission from laser plasma sources in inertial confinement fusion. The existing calibration methods for single reflecting mirrors, crystals, gratings, filters, and X-ray detectors are not applicable for such X-ray microscopes due to the specific optical structure and the restrictions of object-image relation. This article presents a reliable and efficient method that can be performed using a divergent X-ray source and an energy dispersive Si-PIN (silicon positive-intrinsic-negative) detector in an ordinary X-ray laboratory. The transmission theory of X-ray flux in imaging diagnostics is introduced, and the quantities to be measured are defined. The calibration method is verified by a W/Si multilayer-coated Kirkpatrick-Baez microscope with a field of view of ∼95 μm at 17.48 keV. The mirror reflectance curve in the 1D coordinate is drawn with a peak value of 20.9% and an uncertainty of ∼6.0%.