Use of the 22C3 anti–PD-L1 antibody to determine PD-L1 expression in multiple automated immunohistochemistry platforms

作者:Marius Ilié, Shirin Khambata‐Ford, Christiane Copie‐Bergman, Lingkang Huang, Jonathan Juco, Véronique Hofman, Paul Hofman · 发表于:PLoS ONE · 年份:2017 · DOI:10.1371/journal.pone.0183023 · 被引用次数:100 · 研究领域:Cancer Immunotherapy and Biomarkers、Radiomics and Machine Learning in Medical Imaging、Lung Cancer Diagnosis and Treatment

BACKGROUND: For non-small cell lung cancer (NSCLC), treatment with pembrolizumab is limited to patients with tumours expressing PD-L1 assessed by immunohistochemistry (IHC) using the PD-L1 IHC 22C3 pharmDx (Dako, Inc.) companion diagnostic test, on the Dako Autostainer Link 48 (ASL48) platform. Optimised protocols are urgently needed for use of the 22C3 antibody concentrate to test PD-L1 expression on more widely available IHC autostainers. METHODS: We evaluated PD-L1 expression using the 22C3 antibody concentrate in the three main commercially available autostainers Dako ASL48, BenchMark ULTRA (Ventana Medical Systems, Inc.), and Bond-III (Leica Biosystems) and compared the staining results with the PD-L1 IHC 22C3 pharmDx kit on the Dako ASL48 platform. Several technical conditions for laboratory-developed tests (LDTs) were evaluated in tonsil specimens and a training set of three NSCLC samples. Optimised protocols were then validated in 120 NSCLC specimens. RESULTS: Optimised protocols were obtained on both the VENTANA BenchMark ULTRA and Dako ASL48 platforms. Significant expression of PD-L1 was obtained on tissue controls with the Leica Bond-III autostainer when high concentrations of the 22C3 antibody were used. It therefore was not tested on the 120 NSCLC specimens. An almost 100% concordance rate for dichotomized tumour proportion score (TPS) results was observed between TPS ratings using the 22C3 antibody concentrate on the Dako ASL48 and VENTANA BenchMark ULTRA platfo...