Low Contrast Agent and Radiation Dose Protocol for Hepatic Dynamic CT of Thin Adults at 256–Detector Row CT: Effect of Low Tube Voltage and Hybrid Iterative Reconstruction Algorithm on Image Quality

作者:Takeshi Nakaura, Shinichi Nakamura, Natsuki Maruyama, Yoshinori Funama, Kazuo Awai, Kazunori Harada, Shouzaburou Uemura, Yasuyuki Yamashita · 发表于:Radiology · 年份:2012 · DOI:10.1148/radiol.12111082 · 被引用次数:134 · 研究领域:Radiation Dose and Imaging、Advanced X-ray and CT Imaging、Cardiac Imaging and Diagnostics

PURPOSE: To evaluate the effect on image quality of a low contrast agent dose and radiation dose protocol at abdominal dynamic computed tomography (CT) with a low tube voltage, hybrid iterative reconstruction algorithm, and a 256- detector row scanner. MATERIALS AND METHODS: This prospective study received institutional review board approval, and prior informed written consent was obtained from all patients. Seventy-four patients undergoing hepatic dynamic CT were randomly assigned to one of two protocols: Thirty-nine patients underwent scanning with the conventional 120-kVp protocol and the other 35 patients underwent scanning with an 80-kVp tube voltage and a 40% reduction in contrast agent dose. The 80-kVp images were also postprocessed with a hybrid iterative reconstruction algorithm. The estimated effective radiation dose of each protocol was compared and the image noise and contrast-to-noise-ratio (CNR) of the 120-kVp, 80-kVp, and hybrid iterative reconstructed 80-kVp images were evaluated by using the Student t test. RESULTS: The effective radiation dose was 51% lower during the hepatic arterial phase (HAP) and 48% lower during the portal venous phase (PVP) with the 80-kVp protocol than with the 120-kVp protocol (HAP: 5.6 mSv ± 1.0 [standard deviation] vs 11.6 mSv ± 3.3; PVP: 5.8 mSv ± 0.7 vs 11.2 mSv ± 3.2, respectively). The hybrid iterative reconstruction decreased image noise by 23% during the HAP (9.2 ± 1.9 vs 12.0 ± 2.6) and by 24% during the PVP (9.4 ± 1.8 vs 12...