Simultaneous estimation of T 2 and apparent diffusion coefficient in human articular cartilage in vivo with a modified three‐dimensional double echo steady state (DESS) sequence at 3 T

作者:Ernesto Staroswiecki, Kristin L. Granlund, Marcus T. Alley, Garry E. Gold, Brian A. Hargreaves · 发表于:Magnetic Resonance in Medicine · 年份:2011 · DOI:10.1002/mrm.23090 · 被引用次数:93 · 研究领域:Advanced MRI Techniques and Applications、Advanced Neuroimaging Techniques and Applications、Bone and Joint Diseases

T(2) mapping and diffusion-weighted imaging complement morphological imaging for assessing cartilage disease and injury. The double echo steady state sequence has been used for morphological imaging and generates two echoes with markedly different T(2) and diffusion weighting. Modifying the spoiler gradient area and flip angle of the double echo steady state sequence allows greater control of the diffusion weighting of both echoes. Data from two acquisitions with different spoiler gradient areas and flip angles are used to simultaneously estimate the T(2) and apparent diffusion coefficient of each voxel. This method is verified in phantoms and validated in vivo in the knee; estimates from different regions of interest in the phantoms and cartilage are compared to those obtained using standard spin-echo methods. The Pearson correlations were 0.984 for T(2) (∼2% relative difference between spin-echo and double echo steady state estimates) and 0.997 for apparent diffusion coefficient (˜1% relative difference between spin-echo and double echo steady state estimates) for the phantom study and 0.989 for T(2) and 0.987 for apparent diffusion coefficient in regions of interest in the human knee in vivo. High accuracy for simultaneous three-dimensional T(2) and apparent diffusion coefficient measurements are demonstrated, while also providing morphologic three-dimensional images without blurring or distortion in reasonable scan times.