Gas Chromatographic System for Precise, Rapid Analysis of Nitrous Oxide
作者:Arvin R. Mosier, Laura Mack · 发表于:Soil Science Society of America Journal · 年份:1980 · DOI:10.2136/sssaj1980.03615995004400050048x · 被引用次数:205 · 研究领域:Advanced Chemical Sensor Technologies、Analytical Chemistry and Chromatography、Mass Spectrometry Techniques and Applications
Abstract A gas chromatographic (GC) system for the sensitive, rapid, and precise analysis of nitrous oxide (N 2 O) was developed. The system employs a two‐part GC column that can be separated and the first section back‐flushed to remove contaminants while sample analysis continues. Nitrous oxide was detected by an electron capture detector. The GC system was suitable for analyzing small volumes of gas samples for ambient levels of N 2 O at the rate of 10 samples per hour with a precision of about 1 ppb.