Scholay

学术搜索 · AI 审稿 · LaTeX 协作

Silicon-Embedded Receiving Coil for High-Efficiency Wireless Power Transfer to Implantable Biomedical ICs

作者:Rongxiang Wu, Salahuddin Raju, Mansun Chan, J.K.O. Sin, C. Patrick Yue · 发表于:IEEE Electron Device Letters · 年份:2012 · DOI:10.1109/led.2012.2225135 · 被引用次数:50 · 研究领域:Wireless Power Transfer Systems、Energy Harvesting in Wireless Networks、Wireless Body Area Networks

In this letter, a silicon-embedded receiving coil is designed and fabricated for high-efficiency wireless power transfer to implantable biomedical ICs. The 4.5 mm × 4.5 mm embedded receiving coil achieved a large inductance of 4 μH and a high peak quality factor of 20 at 2.8 MHz. Measurement results of an inductive power link using the embedded receiving coil and a conventional printed-circuit-board transmitting coil (2 cm × 2 cm) demonstrated peak voltage gains of 0.84 and 0.24 and peak efficiency values of 30% and 4.3% for separation distances of 5 and 12 mm, respectively. This is the best reported wireless power transmission efficiency for separation distance similar to the implant chip size.