Radiation Dose Reduction to the Male Gonads During MDCT: The Effectiveness of a Lead Shield
作者:Christian Hohl, Andreas Horst Mahnken, Ernst Klotz, Marco Das, A Stargardt, Georg Mühlenbruch, Thorsten Schmidt, Rolf W. Günther, Joachim Ernst Wildberger · 发表于:American Journal of Roentgenology · 年份:2005 · DOI:10.2214/ajr.184.1.01840128 · 被引用次数:72 · 研究领域:Radiation Dose and Imaging、Heavy Metal Exposure and Toxicity、Radiation Shielding Materials Analysis
OBJECTIVE: Our study was designed to quantify the effect of a standard gonad shield on the testicular radiation exposure due to scatter during routine abdominopelvic MDCT. SUBJECTS AND METHODS: Routine abdominopelvic MDCT was performed in 34 patients with gonadal lead shielding and 32 patients without this shielding; the testes were not exposed to the direct beam during the examination. We estimated the testicular dose administered with thermoluminescent dosimetry, taking into account each patient's body weight and body mass index (BMI). RESULTS: With a 1-mm lead shield, the mean testicular dose was reduced from 2.40 to 0.32 mSv, a reduction of 87%. The difference was found to be statistically significant (p < 0.0001). No correlation between testicular dose and body weight or BMI was found. CONCLUSION: Shielding the male gonads reduces the testicular radiation dose during abdominopelvic MDCT significantly and can be recommended for routine use.