CT of the chest: minimal tube current required for good image quality with the least radiation dose.

作者:John Robert Mayo, Thomas E. Hartman, K S Lee, Steven L. Primack, Sverre Vedal, Néstor L. Müller · 发表于:American Journal of Roentgenology · 年份:1995 · DOI:10.2214/ajr.164.3.7863879 · 被引用次数:158 · 研究领域:Radiation Dose and Imaging、Cardiac Imaging and Diagnostics、Advanced X-ray and CT Imaging

OBJECTIVE: We wanted to determine minimal tube current (mAs) required for consistently good image quality on conventional 10-mm collimation chest CT and effect of tube current reduction on detection of mediastinal and lung abnormalities. Tube current reduction is desirable to reduce patient radiation dose. SUBJECTS AND METHODS: Prospectively, 30 consecutive patients (mean weight, 68 kg; range, 34-93 kg) older than 45 undergoing conventional chest CT with standard technique (120 kVp, 400 mAs) had four additional sections imaged at reduced tube current (200, 140, 80, 20 mAs) at two levels (tracheal carina and left atrium). CT scans were evaluated in random order by two independent observers who were blinded to technical factors used. Subjective image quality was graded on a five-point scale from non-diagnostic to excellent. Visualization of mediastinal adenopathy (n = 18), pleural plaques (n = 17), effusions (n = 28), lung parenchymal nodules (n = 37), and emphysema (n = 15) were assessed. The 400 mAs scan was considered the reference standard. RESULTS: When compared with the reference technique (400 mAs), the first and second (200 mAs and 140 mAs) reduction levels showed no significant difference (p > .05) in subjective image quality. A significant difference (p < .001) was seen at the third and fourth (80 mAs and 20 mAs) reduction levels. However, no significant difference (p > .05) was seen in detection of mediastinal or lung parenchymal abnormalities with different tube cur...