Scholay

学术搜索 · AI 审稿 · LaTeX 协作

Certification of Standard Reference Material 660B

作者:David R. Black, Donald Windover, Albert Henins, James J. Filliben, James P. Cline · 发表于:Powder Diffraction · 年份:2011 · DOI:10.1154/1.3591064 · 被引用次数:106 · 研究领域:Nuclear Physics and Applications、X-ray Diffraction in Crystallography、Nuclear Materials and Properties

This report describes SRM 660b, the third generation of this powder diffraction SRM used primarily for determination of the instrument profile function (IPF). It is certified with respect to unit-cell parameter. It consists of approximately 6 g LaB 6 powder prepared using a 11 B isotopically enriched precursor material so as to render the SRM applicable to the neutron diffraction community. The microstructure of the LaB 6 powder was engineered to produce a crystallite size above that where size broadening is typically observed and to minimize the crystallographic defects that lead to strain broadening. A NIST -built diffractometer, incorporating many advanced design features, was used to certify the unit-cell parameter of the LaB 6 powder. Both type A, statistical, and type B, systematic, errors have been assigned to yield a certified value for the unit-cell parameter of a =0.415691(8) nm at 22.5°C.