Evaluation of the accuracy of 3DVH software estimates of dose to virtual ion chamber and film in composite IMRT QA

作者:Arthur J. Olch · 发表于:Medical Physics · 年份:2011 · DOI:10.1118/1.3666771 · 被引用次数:89 · 研究领域:Advanced Radiotherapy Techniques、Radiation Therapy and Dosimetry、Radiation Detection and Scintillator Technologies

PURPOSE: A novel patient-specific intensity modulated radiation therapy (IMRT) QA system, 3DVH software and mapcheck 2, purports to be able to use diode array-measured beam doses and the patient's DICOM RT plan, structure set, and dose files to predict the delivered 3D dose distribution in the patient for comparison to the treatment planning system (TPS) calculated doses. In this study, the composite dose to an ion chamber and film in phantom predicted by the 3DVH and mapcheck 2 system is compared to the actual measured chamber and film doses. If validated in this context, then 3DVH can be used to perform an equivalent dose analysis as that obtained with film dosimetry and ion chamber-based composite IMRT QA. This is important for those losing their ability to perform film dosimetry for true composite IMRT QA and provides a measure of confidence in the accuracy of 3DVH 3D dose calculations which may replace phantom-based IMRT QA. METHODS: The dosimetric results from 15 consecutive patient-specific IMRT QA tests performed by composite field irradiation of ion chamber and EDR2 film in a solid water phantom were compared to the predicted doses for those virtual detectors based on the calculated 3D dose by the 3DVH software using mapcheck 2 measured doses of each beam within each plan. For each of the 15 cases, immediately after performing the ion chamber plus film measurements, the mapcheck 2 was used to measure the dose for each beam of the plan. The dose to the volume of the v...