Lavage Solution Temperature Influences Depth of Chondrocyte Death and Surface Contouring during Thermal Chondroplasty with Temperature-Controlled Monopolar Radiofrequency Energy

作者:Yan Lü, Ryland B. Edwards, Shane J. Nho, Brian J. Cole, Mark D. Markel · 发表于:The American Journal of Sports Medicine · 年份:2002 · DOI:10.1177/03635465020300050601 · 被引用次数:78 · 研究领域:Total Knee Arthroplasty Outcomes、Osteoarthritis Treatment and Mechanisms、Infrared Thermography in Medicine

BACKGROUND: Although radiofrequency energy can smooth and contour cartilage surface, it has deleterious effects on chondrocyte viability. HYPOTHESIS: Monopolar thermal chondroplasty in a 37 degrees C lavage solution, as compared with a 22 degrees lavage solution, will reduce chondrocyte death and result in greater smoothing of the articular cartilage surface. STUDY DESIGN: Controlled laboratory study. METHODS: Sixteen chondromalacic samples from patients undergoing total knee arthroplasty were divided into two groups: 22 degrees C and 37 degrees C lavage solution. Each sample was divided into two equal parts and half of each group was treated for 10 seconds and the other half for 15 seconds. RESULTS: Confocal laser microscopy demonstrated that the depth of chondrocyte death in the 37 degrees C lavage solution group was significantly less (range, 200 to 340 microm) than that in the 22 degrees C solution group for both 10- and 15-second treatment times. Scanning electron microscopy demonstrated that the cartilage surface in the 37 degrees C lavage solution group was smoother than that in the 22 degrees C solution group for the 10-second treatment time. Energy delivery power in the 37 degrees C lavage solution group was significantly lower than in the 22 degrees C solution group for both treatment times. CONCLUSIONS: Thermal chondroplasty with 37 degrees C lavage solution resulted in less depth of chondrocyte death and produced smoother surfaces than with 22 degrees C solution f...